Z-2483-2025: 3M Ranger Blood/Fluid Warming High Flow Set Catalog Numbers 24355, 24370

Recall record Z-2483-2025 for 3M Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2483-2025
Company in FDA record
3M Company
FDA classification date
Sep 18, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Field action to clarify the flow rates and inlet fluid temperatures, along with location of the outlet temperature for Ranger Blood/Fluid Warming High Flow Sets.

Source facts

Action
An Urgent Medical Device Correction letter was sent to all their customers on 4/21/2025 to inform them that Solventum [formerly 3M Company] clarified flow rates related to the inlet fluid temperature. The device instructions for use (IFU) will be updated to reflect these changes. Customers are informed that there have been no changes to the design or composition of these products that prompted this notification. Customers are instructed to continue to use the products as indicated, considering the updated information provided in the letter. If the device is used at a flow rate outside of the updated information provided, there is a potential risk for the device to administer the fluids at a lower temperature than indicated and cause a decrease in patient temperature. Under the dire circumstances of a massive rapid infusion where high flow rates outside of the updated ranges indicated below may be used, there is a potential for hypothermia. Customers with additional questions or need to report an adverse event, are instructed to contact the Solventum [formerly 3M Health Care] Customer Helpline at 1-800-228-3957, option 6.
Initiated
2025-04-10
Root cause
Device Design
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Belgium, Bosnia and Herzegovina, Canada, Canary Islands, China, Costa Rica, Croatia, Czechia, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Korea, Luxembourg, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, UAE, UK, and Vietnam.
Recall number
Z-2483-2025
Recall status
Open, Classified
CDRH record ID
213764
FDA classified
2025-09-18
Recall event number
96841

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗