Skip to content

Public FDA record

Z-2484-2015: OmniPod Insulin Management System. Only the OmniPod device is affected. Catalog Number(s): POD-ZXP420 (US

Recall record Z-2484-2015 for Insulet Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2484-2015
Company in FDA record
Insulet Corporation
FDA classification date
Sep 4, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

OmniPods (Pods) have a higher rate of failure causing: Cannula fails to deploy/retract causing insulin to be pumped or the audible alarm is displayed on the PDM and pod will not deliver insulin

Source facts

Action
The firm, Insulet Corporation, sent an "URGENT: Voluntary Device Recall" letter dated 7/13/2015 to each consignee/customer via Email notification (where possible) on 7/13/15 as well as Federal Express. Customers not responding to the email or Federal Express will receive follow up phone calls. The letter described the product, problem and actions to e taken. The customers were instructed to do the following: If you have Pods from any of these lots, set them aside, we will replace them at no charge; arrange for return and replacement by choosing one of the following methods: 1. Complete the online form at https://www.myomnipod.com/podinfo (fastest method). 2. Return the enclosed reply card by mail or fax to 1-855-407-3729. 3. Contact Insulet Customer Care by calling 1-855-407-3729 (any time day or night). If you do not have any unused Pods form these lots, reply using one of the methods above. If you have any question, contact Acting VP RA/QA at 978-600-7000. EXPANDED RECALL: Insulet Corp. issued on 11/2/15 the URGENT: Field Safety Notification via Email notification and Federal Express. Letter describes problem of that certain lots of the pod's needle failed to deploy or there is a delay in the deployment of the needle. Customers not responding to the email or Federal Express will receive additional mailing and/or follow up phone calls. Call Customer Care at 1-855-407-3729 if you have any questions regarding this Field Safety Notification
Initiated
2015-07-13
Root cause
Nonconforming Material/Component
Terminated
2017-01-26
Distribution
Worldwide Distribution: US (Nationwide) and country of: Switzerland.
Recall number
Z-2484-2015
Recall status
Terminated
CDRH record ID
138558
FDA classified
2015-09-04
Recall event number
71651

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.