Skip to content

Public FDA record

Z-2484-2019: HS Hi-R NEO 900A NIR Reference # 657 822 Serial # from 101 Product Usage: A wide range of surgical inter

Recall record Z-2484-2019 for Haag-Streit USA Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2484-2019
Company in FDA record
Haag-Streit USA Inc
FDA record creation date
Sep 5, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

software error in the central control unit of the floor stand.

Source facts

Action
1. Do not perform any other functions at the same time when operation the XY control, such as changes in the light intensity. 2. Perform a software update of the floor stand. you will receive the software version 11.4 through your local representative which will fix the bug. 3. please send us the enclosed response form filled In as soon as possible. 4. If you have passed the products to third parties, please forward a copy of this information and inform the below mentioned contact person. The software update will be organized through Haag-Streit USA. For questions concerning this procedure, please contact: Brian Gallagher Phone: (513) 486-1522 Email: fieldsafetynotice@haag-streit-usa .com
Initiated
2019-07-31
Root cause
Software Manufacturing/Software Deployment
Terminated
2020-04-16
Distribution
US in the state of OH
Recall number
Z-2484-2019
Recall status
Terminated
CDRH record ID
175756
FDA record created
2019-09-05
Recall event number
83571

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗