Z-2485-2023: Wireless Foot Switch distributed with Philips Allura Xper and Azurion Interventional Fluoroscopic X-Ray S

Recall record Z-2485-2023 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2485-2023
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA classification date
Aug 30, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Loss of availability of the wireless foot switch during procedures.

Source facts

Action
Customers were sent an URGENT Medical Device Correction letter dated 6/23/2023 via USPS Certified Mail. Customers are instructed to ensure that the wired footswitch is always connected to the system in case the wireless foot switch becomes unavailable. Customers are instructed to download a copy of the addendum to the IFU for the wireless foot switch and the Quick Reference Card at Http://philips.com/doc_library, and distribute these materials to those who use the system and retain a copy with the system's IFU. These documents are to be placed in an area where they are likely to be seen/viewed with system documentation. Customers are to return the attached response form to Philips no later than 30 days from receipt of the notification to IGT_Recalls@philips.com. Philips will contact customers to schedule a Field Service Engineer visit to confirm that the wired foot switch is connected to affected systems, perform a check of the charger, and provide customers with a hardcopy of the addendum. Customers with questions can contact Philips Technical Support Line at 1-800-722-9377.
Initiated
2023-06-30
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Bosnia and Herzegovina, Brazil, Chile, China, Croatia, Czech Republic, Denmark, Egypt, Fiji, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of, Kosovo, Latvia, Lebanon, Libya, Malaysia, Maldives, Malta, Martinique, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, State of, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, and Yemen.
Recall number
Z-2485-2023
Recall status
Open, Classified
CDRH record ID
202568
FDA classified
2023-08-30
Recall event number
92782

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗