Z-2486-2020: PERMA-HAND/MERSILK SILK BLU and VIRGIN SILK BLU, Product Codes EH7457G EH7983G FV768 FV769 FV7771 U7058

Recall record Z-2486-2020 for Ethicon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2486-2020
Company in FDA record
Ethicon, Inc.
FDA record creation date
Jul 2, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Lots of size 8-0 Blue twisted Silk Suture may not meet the stringent full shelf life of 5 years

Source facts

Action
There is no US impact. The firm began notification of customers on June 10, 2020 via URGENT FIELD SAFETY NOTIFICATION letters. Customers were informed that the shelf life of the product is being reduced from 5 years to 3 years (36 months). Any 8-0 Blue Twisted Silk Suture product received prior to June 2019 labeled with an expiration date of 2024-06-20 or earlier is affected by this Field Safety Notification. Customers were asked to acknowledge the notification by completing and returning the business reply form included with the mailing. If you have additional questions regarding this Field Safety Notification or to report any customer complaints, please contact Ethicon Customer Support Center at 1-877-ETHICON (1-877-384-4266). The Customer Support Center is open Monday through Friday, 8:00 AM to 6:00 PM ET. ***Updated 6/29/20*** On 6/12/2020, the recalling firm updated the field action from "Notification" to "Removal". All affected devices are now required to be returned.
Initiated
2020-06-10
Root cause
Nonconforming Material/Component
Distribution
No domestic distribution. Foreign distribution to Algeria Angola Australia Austria Azerbaijan Bahrain Belgium Bosnia and Herzegovina Canada Chad Chile Czechia Egypt Finland France Gabon Germany Greece Hong Kong Iceland India Iran Ireland Israel Italy Japan Jordan Kenya Kuwait Latvia Lebanon Libya Lithuania Mauritius Morocco Mozambique Netherlands (the) New Caledonia New Zealand Nigeria Oman Philippines Poland Portugal Qatar Republic of North Macedonia Russian Federation (the) Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sweden Switzerland Thailand Tunisia Turkey United Arab Emirates (the) United Kingdom of Great Britain and Northern Ireland (the)
Recall number
Z-2486-2020
Recall status
Open, Classified
CDRH record ID
181837
FDA record created
2020-07-02
Recall event number
85820

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗