Open original FDA record ↗Recall record
- FDA identifier
- Z-2487-2012
- Company in FDA record
- Stryker Howmedica Osteonics Corp.
- FDA classification date
- Sep 28, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Stryker Orthopaedics has become aware of a burr, or a metal shaving, observed in the internal threaded portion of the Cone Body Trial. The typical size of the shaving observed was 1-2 mm in length and 0.037 mm squared in cross section.
Source facts
- Action
- Stryker Orthopaedics sent a Urgent Product Field Action - Product Inspection Branch Acknowledgement Form on April 2, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. All inspections must be completed immediately as per the provided Product Correction Bulletin. Please document the results of your inspection on the Tracking Log, indicating conforming devices as "Not Affected" and non-conforming devices as "To be returned." If the devices listed on your tracking log have previously been returned or are otherwise not on hand, indicate the appropriate disposition following our usual process. Please assist us in meeting our regulatory obligation by faxing back the attached Product Acknowledgment Form and Tracking Log within 10 business days to 855-251-3635. Any further questions please call 201-831-5523.
- Initiated
- 2012-04-02
- Root cause
- Device Design
- Terminated
- 2013-06-13
- Distribution
- Nationwide distribution
- Recall number
- Z-2487-2012
- Recall status
- Terminated
- CDRH record ID
- 111117
- FDA classified
- 2012-09-28
- Recall event number
- 62594