Z-2488-2021: Phantom Nail System, Drill, 4.6 x 300mm, Cannulated, 3/16" Sq. Connection p28 P99-110-4630

Recall record Z-2488-2021 for Paragon 28, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2488-2021
Company in FDA record
Paragon 28, Inc.
FDA record creation date
Sep 17, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A thin wall condition between the inner cannulation and the flutes of the drill which potentially could cause a fracture through the wall of the drill to the cannulation beneath.

Source facts

Action
On 08/12/2021, the Recalling Firm emailed to its distributors an "Medical Device Recall" Notification Letter inform them that it was identified that there is a thin wall condition between the inner cannulation and the flutes of the drill. At worst case, the defect is a fracture through the wall of the drill to the cannulation beneath. This feature is potentially nonconforming in some units within the lot identified below. The distributor was instructed to: 1) Immediately check its inventory and hospitals/health care facility within their region and to identify, discontinue use, quarantine and not distribute any further the affected products. 2) Complete the Medical Device Recall Response Form and return to mkinnee@paragon28.com. The form is to be completed and returned even if no physical product is in their possession. 3) Immediately return affected devices along with a copy of the Medical Device Recall Response Form to the Recalling Firm at Paragon28, Attn: Recalls, RN-008, 14445 Grasslands Dr., Englewood, CO 80112. For questions or further assistance - contact the Recalling Firm at: email - mkinnee@paragon28.com or complaints@paragon28.com Phone - 949-280-7688
Initiated
2021-07-30
Root cause
Process change control
Terminated
2021-12-15
Distribution
U.S. Nationwide distribution in the states of CA, FL, GA, ID, KS, KY, LA, MI, MO, NC, NY, TX, VA, WA and WV. O.U.S.: N/A
Recall number
Z-2488-2021
Recall status
Terminated
CDRH record ID
188985
FDA record created
2021-09-17
Recall event number
88553

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗