Z-2489-2023: Upper roller bracket (part number 1088370) for the Gen2 gantry enclosure for the Radixact Treatment Deliv

Recall record Z-2489-2023 for Accuray Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2489-2023
Company in FDA record
Accuray Incorporated
FDA classification date
Aug 31, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It is possible the upper roller bracket in the Radixact System gantry enclosure may not provide enough clearance for service personnel to use a jib crane for moving components while servicing the system, resulting in possible injury to the service personnel.

Source facts

Action
The recalling firm began issuing letters dated 7/19/2023 via hand-delivery by the field service engineer. The letter explained the issue and explained that if their Radixact System is being serviced by a third-party provider, please ensure they are also aware and take relevant precautions while servicing the system. The product impacted by the issue was detailed. The letter explained the issue arises only during the servicing of the Radixact System and customers can continue the clinical use of their system. The letter informs the consignee that Accuray will inspect the upper roller bracket of each impacted Radixact system to confirm there is adequate clearance for a jib crane. If there is sufficient clearance, then no action is necessary. If there is insufficient clearance, then the old bracket will be replaced with the new bracket following the released updated service procedures. An acknowledgment form was enclosed documenting the consignee confirms understanding the content of the field safety notice and has distributed the information to all applicable members of their staff.
Initiated
2023-08-03
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of CA, GA, IL, LA, NJ, OR, PA, SC, SD, TX, and WA. There was government distribution and no military distribution. The countries of Australia, Belgium, Canada, Colombia, France, India, Indonesia, Italy, Japan, Kazakhstan, Malaysia, Poland, Saudi Arabia, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.
Recall number
Z-2489-2023
Recall status
Completed
CDRH record ID
202523
FDA classified
2023-08-31
Recall event number
92772

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗