Randox Laboratories realigned RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target for the Randox Calibration Serum Level 3 (for calibrator lot 961UE onwards).
Source facts
Action
Randox Laboratories Ltd. issued Urgent Medical Device Correction letter on 7/12/19 advising users of the problem, health risk and action to take: Replace all Quality Control value sheets with the revised sheets available on www.randox.com . Discuss the contents of this notice with your Medical Director. Complete and return the response form 12187-QA to technical.services@randox.com within five working days.