Open original FDA record ↗Recall record
- FDA identifier
- Z-2495-2019
- Company in FDA record
- Philips North America, LLC
- FDA record creation date
- Sep 9, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The thickness of the connector to the trunk cable on affected limb lead sets may be greater than specified. which may prevent or impede the performance of I 2-Lead ECG measurements.
Source facts
- Action
- Philips sent a Customer Information letter dated July 2019 to affected customers. The letter identified the affected product, problem and actions to betaken. Customers were instructed to: 1. To identify all affected limb lead sets. 2. To fill-out, sign and send a reply form that is provided. 3. To discard affected limb lead sets following approved medical waste disposal methods per facility and/or local regulations. 4. Do not return any affected product to Philips if you need any further information or support concerning this issue, please contact your local Philips representative.
- Initiated
- 2019-07-19
- Root cause
- Nonconforming Material/Component
- Terminated
- 2020-06-03
- Distribution
- Worldwide Distribution - US Nationwide State KY AZ CA UT OH NY SC CO FL IN TX ND KS GA Canada Germany
- Recall number
- Z-2495-2019
- Recall status
- Terminated
- CDRH record ID
- 175736
- FDA record created
- 2019-09-09
- Recall event number
- 83560