Z-2496-2020: TOMTEC-ARENA TTA2;software version no. 2.20 and lower; Picture archiving and communications system. Soft

Recall record Z-2496-2020 for Tomtec Imaging Systems Gmbh. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2496-2020
Company in FDA record
Tomtec Imaging Systems Gmbh
FDA record creation date
Jul 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm discovered a software issue associated with the Image-Com module/clinical application package (CAP) interface for 3D application in the ARENA software, where measurements from one study might be stored to another study.

Source facts

Action
Customers were sent the Urgent Medical Device Correction, Field Safety Notice letters, issued on 09/23/2019 via certified mail (USPS) with return receipt. The firm asked customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction Letter. The instructions describe how to delete all measurements if a Clinical Application Package (CAP) has been initiated and the clip/image is rejected. The instructions to the customers to avoid the software issue are as follows: If a CAP has been initiated and the clip/image is rejected and one of the messages described under problem description is displayed, take the following actions: 1. Close the message dialog from the CAP and move back to IMAGE-COM 2. Press the Worksheet Button 3. Navigate through the measurements menu and check all packages (4D LVA, 4D MVA, 4D RVF, 2D CPA, 4D CV) in all groups in the exam types Adult Echo and Pediatric echo. Definition of hierarchy: -Exam Type (e.g. Adult Echo, Pediatric Echo, ...) Group (eg: Left Ventricle, Mitral Valve, Right Ventricle, Left Ventricle) - Package (e.g.: 4D LVA, 4D MVA, 4D RVF, 2D CPA, 4D CV) - SubPackage (optional) o Measurement 4. Once found, delete all measurements on the above described packages by using the available delete functions. 5. All packages described above in the worksheet shall be completely empty. 6. Once finished, close the Worksheet. If you suspect that the patient data software issue has occurred, contact the local representative for support. Complete the reply form and return as soon as possible via email to support@tomtecusa.com. The firm will resolve the issue by providing a software update, at no cost, to correct the issue in the software version that is currently installed on your system. A representative will contact to schedule the software installation. If further information or support is needed concerning this issue, contact the local representative at
Initiated
2019-09-23
Root cause
Software design
Distribution
US:AL, AZ, CA, CT, IL, MA, MI,NC, NJ, NY, OH, PA, RI, TX, VA. OUS: Australia, Canada, Czech Republic, France, Germany, Italy, Latvia, Sweden, Switzerland, United Kingdom.
Recall number
Z-2496-2020
Recall status
Open, Classified
CDRH record ID
180732
FDA record created
2020-07-04
Recall event number
85392

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗