Open original FDA record ↗Recall record
- FDA identifier
- Z-2497-2020
- Company in FDA record
- Ra Medical Systems, Inc.
- FDA classification date
- Jul 4, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A software issue was identified which could result in user or patient injury, or may adversely impact laser performance. It is possible that an unintended release of laser light (radiation) may occur, injuring the user or patient.
Source facts
- Action
- The Urgent Medical Device Correction notification was sent as certified mail USPS letter to consignees on 1/31/2020. The requested actions are as follows: 1. Do not use the excimer laser until a service technician has installed the new version of software to correct the issue. 2. Complete and return the acknowledgement form, whether you have any of the affected product or not, indicating: * Whether you have the excimer laser in question at your facility, *If you do have the excimer laser, the physical address where the product is located; and, *The person to be contacted to schedule a service call to correct the issue. Should you have any questions or need further assistance call (442) 325-5044.
- Initiated
- 2020-01-23
- Root cause
- Software design
- Terminated
- 2021-11-01
- Distribution
- US: AL, AZ, CA, FL, GA, ID, IL, IN, KS, KY, LA, MI, MD, NC, NV, NY, OK, OR, PA, TN,TX, WI. OUS: Italy and Japan.
- Recall number
- Z-2497-2020
- Recall status
- Terminated
- CDRH record ID
- 181788
- FDA classified
- 2020-07-04
- Recall event number
- 85790