A gluing step was not performed during the manufacturing of the affected device's insertion tube. There is a potential for the insertion tube to develop a leak, which poses an infection control risk.
Source facts
Action
On August 17, 2021, the firm issued Urgent Medical Device Correction letters to affected customers.
Customers were informed that Olympus will repair your unit free of charge. An Olympus representative will contact you to make arrangements for return of your affected bronchoscope.
If you have further questions about this recall, please contact laura.storms@olympus.com for additional information.
Initiated
2021-08-17
Root cause
Under Investigation by firm
Terminated
2024-01-12
Distribution
US Nationwide distribution in the states of CA, IL, MA, MD, MI, NY, OH, OR, and PA.
Recall number
Z-2497-2021
Recall status
Terminated
CDRH record ID
188993
FDA record created
2021-09-21
Recall event number
88560