Open original FDA record ↗Recall record
- FDA identifier
- Z-2500-2018
- Company in FDA record
- Philips Healthcare
- FDA record creation date
- Jul 25, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination.
Source facts
- Action
- An Urgent Medical Device Correction letter dated June 11, 2018 was sent to customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers, until a software revision that corrects this issue becomes available, users should verify that the desired shutter position is set when performing the run after the first time the procedure type is changed during a single examination. This can be accomplished by first selecting a different procedure and then reselect the original procedure on the Xper Module or on the Data Monitor. Customer shall ensure that all staff with access to the affected systems are informed of the contents of this Field Safety Notice. A copy of this Field Safety Notice shall be placed together with the documentation of the system until the system has been corrected by Philips. For questions contact your local Philips representative Technical Support Line 1-800-722-9377.
- Initiated
- 2018-06-11
- Root cause
- Software design
- Terminated
- 2024-07-03
- Distribution
- US Nationwide Distribution
- Recall number
- Z-2500-2018
- Recall status
- Terminated
- CDRH record ID
- 165744
- FDA record created
- 2018-07-25
- Recall event number
- 80402