Z-2502-2018: VOLISTA StandOP Surgical Light, ARD568811961 VLT600DF AIM STP; ARD568821961 VLT600DF AIM STP QL; ARD5688

Recall record Z-2502-2018 for Maquet SAS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2502-2018
Company in FDA record
Maquet SAS
FDA record creation date
Jul 19, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket breaking, causing the light head to detach from the arm and become suspended only from the cabling inside the arm. Use of an affected device may result in serious injury to patient and/or medical staff if the light head detaches and drops, impacting patient and/or medical staff during demonstration, service, or use.

Source facts

Action
On June 14, 2018, the firm issued an Urgent Medical Device Field Correction letter via FedEx to affected customers. Customers were advised of the potential light head detachment on VOLISTA StandOP Surgical Lights. Customers were instructed to do the following: 1. Ensure that all users were made aware of the recall; distributors were asked to forward the recall letter to their customers' attention for appropriate action. 2. Check your VOLISTA StandOP serial number versus the list of affected devices on page 5 and ensure it is available for a Getinge Service technician visit. 3. Per the VOLISTA Instructions for Use*, Section 11.2 (General Maintenance) - Daily Checks (performed by the user), examine your affected VOLISTA StandOP surgical light immediately. If the bracket in the device is improperly assembled, mechanical play will be observed when the light head is manipulated in a way that is not expected to be encountered in clinical use. Further instructions for testing the device are included in the letter. 4. Complete and sign the enclosed Customer Response Form (Page 5) and return to Getinge using the fax number or email address provided on the form. A Getinge Service person will contact the person you listed on the Customer Response Form to schedule service for an update of your device, free of charge. You can also arrange a visit by a Getinge-trained or authorized service technician by calling Getinge Customer Support at (800) 950-9912 (select option 1). Customer Service hours: 8:00 AM and 5:00 PM (EST) Monday to Friday.
Initiated
2018-06-14
Root cause
Process control
Terminated
2022-06-23
Distribution
Worldwide Distribution - US Nationwide in the states of AR, AZ, CA, CO, CT, FL, GA, IA, ID, KS, KY, LA, MD, ME, MO, NE, NY, OH, OR, TN, TX, UT, and WA. Foreign distribution to Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil Cameroon, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Ghana, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Korea, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Namibia, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Poland, Portugal, South Korea, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Tanzania, Yemen, and Zimbabwe.
Recall number
Z-2502-2018
Recall status
Terminated
CDRH record ID
165518
FDA record created
2018-07-19
Recall event number
80452

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗