There was a clerical error in order entry which resulted in the recalling firm recieving spring components from a supplier made of "music wire" The recalling frim conducted tests to determine the effects of theis change in materials. The testing showed that the wire would oxidize and there would be early failure of the device.
Source facts
Action
The firm send URGENT MEDICAL DEVICE RECALL NOTICES to its distributors on 7/28/2010. The notices instructed the distributors to discontinue use of the product and return it to Biomet.
Initiated
2010-07-28
Root cause
Employee error
Terminated
2012-10-01
Distribution
US, Australia, Canada, Mexico, Chile, Japan.
Recall number
Z-2504-2010
Recall status
Terminated
CDRH record ID
94040
FDA classified
2010-09-25
Recall event number
56591