Z-2505-2010: Modular Microplasty Cup impactor 1/4" - 28 thread insert , REF 31-400605

Recall record Z-2505-2010 for Biomet, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2505-2010
Company in FDA record
Biomet, Inc.
FDA classification date
Sep 25, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

There was a clerical error in order entry which resulted in the recalling firm recieving spring components from a supplier made of "music wire" The recalling frim conducted tests to determine the effects of theis change in materials. The testing showed that the wire would oxidize and there would be early failure of the device.

Source facts

Action
The firm send URGENT MEDICAL DEVICE RECALL NOTICES to its distributors on 7/28/2010. The notices instructed the distributors to discontinue use of the product and return it to Biomet.
Initiated
2010-07-28
Root cause
Employee error
Terminated
2012-10-01
Distribution
US, Australia, Canada, Mexico, Chile, Japan.
Recall number
Z-2505-2010
Recall status
Terminated
CDRH record ID
94041
FDA classified
2010-09-25
Recall event number
56591

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗