Z-2507-2023: Proclaim Plus 5 Implantable Pulse Generator REF 3670 Product Description: The implantable pulse generato

Recall record Z-2507-2023 for Abbott Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2507-2023
Company in FDA record
Abbott Medical
FDA classification date
Sep 11, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).

Source facts

Action
On 07/20/2023, the firm will send an "URGENT MEDICAL DEVICE CORRECTION" Letter via mail with delivery tracking to customers informing them, Abbott has received complaints from patients who are unable to exit MRI mode, as their Patient Controller has lost the ability to connect or communicate with their IPG while in MRI mode. Situations where this has occurred include where the user deleted the Bluetooth pairing, lost or disabled their Patient Controller, or upgraded the iOS! software on their Patient Controller while in MRI mode. Next Steps for Customers: Per Abbott's Instructions for Use, patients are advised to contact their physician before having an MRI to discuss all critical information regarding MRI scans and MRI mode. In alignment with the Clinician Programmer and Patient Controller Instructions for Use, Abbott recommends physicians do the following: " For the Patient Controller, advise the patient not to delete the paired Bluetooth connection between their IPG and the Patient Controller and not to alter, damage or lose their Patient Controller while the IPG is in MRI mode. " For the Patient Controller prior to entering MRI mode, ensure patients have upgraded their Patient Controller to the latest "Patient Controller NR - US" application from the Apple! App Store!. This version of the application provides instructions for the user not to delete the IPG pairing while the system is in MRI mode. " For the Clinician Programmer, maintain the paired Bluetooth connection between the IPG and the Clinician Programmer by avoiding deleting the pairing and disabling automatic iOS! software upgrades from the iOS! settings. " For questions about this issue or to report patients who are unable to exit MRI mode, please contact your local Abbott representative or Abbott Technical Support at 1-800-727-7846 (option 3) for assistance.
Initiated
2023-07-18
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Korean, Kuwait, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom.
Recall number
Z-2507-2023
Recall status
Open, Classified
CDRH record ID
202212
FDA classified
2023-09-11
Recall event number
92688

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗