Open original FDA record ↗Recall record
- FDA identifier
- Z-2509-2026
- Company in FDA record
- Staar Surgical AG
- FDA classification date
- Jun 18, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a software coding issues for calculating the implant orientation diagram printout for the implantation for toric implantable collamer lenses.
Source facts
- Action
- On 04/20/2026, the firm sent via FedEx an "URGENT: MEDICAL DEVICE RECALL" Letter informing customers that STAAR has confirmed that during that period, the printing of the IOD within the Stella Patient View, Order Confirmation, or Shopping Cart, when performed locally (i.e. by doctor/clinic), was under certain circumstances producing an incorrect diagram. The incorrect diagram was produced in certain circumstances when users created multiple Toric lens line items with different axis values on the same patient eye, such as with backup or secondary lens reservations. The root cause has been identified as a software coding error in Stella 2.0 that did not assign a unique identifier to each reservation under a patient s eye. STAAR has modified the software code to assign a unique identifier, thereby preventing this issue when multiple reservations exist. Any IOD printed from Stella 2.0 after 5:30pm PDT on April 16, 2026 would not have been affected by this issue. For Customer's awareness: We are informing you so that a proper clinical decision can be made regarding patient management and follow-up, based on your assessment of the clinical outcome. Please immediately complete and return the Recall Response Form acknowledging receipt of this recall and report any adverse events associated with this event. If a Toric EVO/EVO+ device were implanted using an incorrect IOD printout, there is the potential for the following adverse events: " Refractive surprise " Blurred vision " Glares/Halos " Other visual disturbances Repositioning or removal of the lens may need to be considered in these cases. For Questions - contact STAAR Sales Representative or Customer Service at customerservice@staar.com or by phone at 1-800-352-7842 Monday through Friday, 6:00am to 5:00pm PDT.
- Initiated
- 2026-04-20
- Root cause
- Software design
- Distribution
- Worldwide - US Nationwide distribution including in the states and territory of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Austria, Australia, Belgium, Canada, Switzerland, Czech Republic, Germany Spain, Finland, France, Great Britain, Hong Kong, Ireland, Israel, India, Iran, Italy, Jordan, Korea, Kuwait, Kazakhstan, Morocco, Malaysia, Netherlands, Oman, Philippines, Poland, Saudi Arabia, Sweden, Singapore, Thailand, Turkey, Taiwan, Ukraine, USA, Vietnam.
- Recall number
- Z-2509-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220499
- FDA classified
- 2026-06-18
- Recall event number
- 99007