Z-2510-2024: IntelliVue Patient Monitor MX500 with L.x and M.x; K.x provided with L.x entitlements Model Number: 86606

Recall record Z-2510-2024 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2510-2024
Company in FDA record
Philips North America Llc
FDA classification date
Aug 5, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Philips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the process, the entitlements of software options MOS, M06, and M20 were removed for software versions L.x and M.x. As a result of these changes Patient Monitors manufactured with or updated to the latest versions of software L.x or M.x will not offer the capabilities> Mmonitors with software version K.x may experience this issue too, due to software version K being out of support and the devices being provided with L.x entitlements offered by software options MOS, M06, and M20. Potential for incorrect or delayed treatment for the patient,

Source facts

Action
Philips North America issued URGENT Medical Device Correction on 6/25/24 to End-Users. Philips Market Organizations will distribute the letters outside of the U.S. Letter states reason for recall, health risk and action to take: Use only default measurement capabilities, as indicated in Table 1 above, for monitoring purposes. Pass this notice to all those who need to be aware within your organization or to any organization where affected devices have been potentially transferred. A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will reload the current device software to enable missing options (MOS, M06, M20). If you need any further information, please contact your local Philips representative or call 1-800-722-9377.
Initiated
2024-06-28
Root cause
Software design
Distribution
Nationwide Foreign: Australia Austria Bangladesh Brazil Canada Chile China Colombia Denmark France Germany Hong Kong India Indonesia Italy Japan Netherlands Portugal Singapore South Africa South Korea Spain Sri Lanka Sweden Switzerland Taiwan Thailand United Kingdom Vietnam
Recall number
Z-2510-2024
Recall status
Open, Classified
CDRH record ID
208808
FDA classified
2024-08-05
Recall event number
94919

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗