Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA classification date
Jul 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
There is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension of some device features (i.e., battery measurements, Capture Management", Atrial Lead Position Check", EffectivCRT" algorithms, and AdaptivCRT"). This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or presence of a magnet.
Source facts
Action
The firm initiated their medical device correction on 06/04/2020 by letter. The letter explained the problem and informed the consignee of the available software update (version 3.2.01) along with download instructions.
Initiated
2020-06-04
Root cause
Software design
Distribution
worldwide
Recall number
Z-2512-2020
Recall status
Open, Classified
CDRH record ID
181828
FDA classified
2020-07-10
Recall event number
85816