Z-2512-2023: PHOROPTOR VRX DIGITAL REFRACTION SYSTEM-Digital refractor intended for the subjective measurement of the

Recall record Z-2512-2023 for Reichert, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2512-2023
Company in FDA record
Reichert, Inc.
FDA classification date
Aug 31, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Phoroptor VRx Head may detach from the bracket that connects the Phoroptor VRx Head to an ophthalmic stand. If the Phoroptor VRx Head detaches from the bracket, it can fall onto a patient or device operator

Source facts

Action
UPDATE July 2024: Reichert issued URGENT MEDICAL DEVICE CORRECTION PROMPT ATTENTION REQUIRED to US distributors and direct account holders via email and Fed'x on 7/3/24 about the extended recall and requested that they return end customer location information to us. Please contact our Reichert Technical Support Team at 1-888-849-8955 or at reichert.support@ametek.com immediately to schedule an appointment with the Reichert Service Team. A Reichert Service Team Member will visit your location, inspect your unit(s), and make any necessary repairs. Below is the Return Merchandise Authorization (RMA) that has been assigned to your unit. Reichert will inspect and repair all units in the relevant manufacturing range. Please schedule an appointment with Reichert even if your unit(s) does not display the defect condition described above. It is necessary that ALL units be inspected by Reichert. If you prefer, you may send your unit(s) to Reichert for inspection and repair. The Reichert Technical Support Team can provide a loaner unit to you using the Return Merchandise Authorization shown above. Inspections, repairs and the loaner unit will be provided at no cost to you. __________________________________________________________________ Reichert contacted authorized distributors and direct accounts who were shipped devices within the scope of the campaign on, July 20, 2023. The letter informed them of the product correction and the need to provide Reichert with the end-user location and contact information for all affected units. The Customer Urgent Medical Device Correction - Prompt Attention Required Letter issued July 27, 2023. Letter states reason for recall, health risk and action to take: While Reichert plans a thorough inspection of all units manufactured between December 2021 and January 2022, pending completion of the Reichert visit to your location, please conduct the abbreviated inspection shown in the image below. If a unit displays the condition
Initiated
2023-07-20
Root cause
Device Design
Distribution
US Nationwide distribution.
Recall number
Z-2512-2023
Recall status
Open, Classified
CDRH record ID
202626
FDA classified
2023-08-31
Recall event number
92805

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗