Z-2512-2024: Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300

Recall record Z-2512-2024 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2512-2024
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Aug 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport.

Source facts

Action
Baxter issued issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 06/28/2024 via USPS. The notice explained the problem with the device, hazard involved, and requested the following: Actions to be Taken by Customers 1. Immediately cease all use of the impacted units until a correction has been completed. If you have alternative tables available, please use them until the service kit is installed by the hospital technician, a Baxter technician, or an authorized representative. 2. Below are two options for correction of the impacted units: a. Self-installation of the service kit (part number 2086594) executed by a hospital-defined technician per the technical service bulletin included with the service kit. Once the self-installation is successful, the impacted units are ready for use. For any questions regarding the installation of the service kit, please contact Baxter Technical Support at 800-445-3720 (select option 2, then option 7) between 8:00 am and 6:00 pm Eastern Time, Monday through Friday. b. A Baxter representative will contact you to schedule inspection of self-corrected units or correct the impacted units at your facility. Do not use the impacted units until the correction has been executed. To request correction of your impacted tables, contact Baxter Technical Support at 800-445-3720 (select option 2, then option 7) between 8:00 am and 6:00 pm Eastern Time, Monday through Friday. 3. Please provide this information to all users of the PST 500 U Precision Surgical Table. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them. 4. If you received this communication directly from Baxter, please acknowledge receipt of this notification by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com, even if you do not have any remaining inventory. Log in to the portal using the account number listed in the enclosed reply
Initiated
2024-06-28
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Recall number
Z-2512-2024
Recall status
Open, Classified
CDRH record ID
208860
FDA classified
2024-08-06
Recall event number
94944

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗