Open original FDA record ↗Recall record
- FDA identifier
- Z-2514-2025
- Company in FDA record
- Preat Corp
- FDA classification date
- Sep 3, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the abutment in the correct orientation, resulting in restoration delay.
Source facts
- Action
- On 06/25/2025, recall notices were mailed to customers who were asked to do the following: 1) Stop/cease use of the product in the scope of this recall 2) This product shall be discarded or returned to the recalling firm 2) All personnel within your company should be made aware of the issue. Any customers to which this product has been further distributed should be notified be provided a copy of this notification and informed to notify their customers. 3) Complete and return the Customer Response Form via email to mgonzales@younginnovations.com If you have any questions or concerns, please call/email the firm at: (800) 232-7732, mgonzales@younginnovations.com
- Initiated
- 2025-06-25
- Root cause
- Process control
- Distribution
- US Nationwide distribution in the states of NY, FL, PA, CA, OH.
- Recall number
- Z-2514-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214944
- FDA classified
- 2025-09-03
- Recall event number
- 97239