Skip to content

Public FDA record

Z-2514-2026: BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to

Recall record Z-2514-2026 for BD SWITZERLAND SARL. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2514-2026
Company in FDA record
BD SWITZERLAND SARL
FDA classification date
Jun 18, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable.

Source facts

Action
On 05/21/2026, the firm sent via email and FedEx an "URGENT: Medical Device Correction" Letter informing customers that, BD has identified an observed issue in how Port C of the BD Connecta Stopcocks operates during use. Reports received indicate that when using polypropylene plastic luer lock syringes, the connection at Port C may over thread the syringe and continue to spin. This behavior has led some clinical users to perceive the connection as unstable. Actions for Clinical Users: -Per current Instructions for Use: "Do not overtighten connections. Note: Resistance between the male luer connector and the BD Connecta Stopcock when rotating indicates the connection is fully engaged and secure. Instruction to follow: 1.Circulate this notice to all those who need to be aware within your organization or to any organization where the potentially affected products have been transferred. 2.Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification. 3.Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787 4.Report any complaints experienced with the use of this product to BD via the North American Regional Complaint Center: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted Mon Fri 8:00am and 5:00pm CT Email: productcomplaints@bd.com For Questions: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com
Initiated
2026-05-21
Root cause
Device Design
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. O.U.S.: Japan
Recall number
Z-2514-2026
Recall status
Open, Classified
CDRH record ID
220006
FDA classified
2026-06-18
Recall event number
98855

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗