Z-2515-2025: Platelia Toxo IgM Catalog Number 26211 UDI Code: 03610520005552 The Platelia Toxo IgM assay is a semi-qu

Recall record Z-2515-2025 for Bio-Rad Laboratories, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2515-2025
Company in FDA record
Bio-Rad Laboratories, Inc.
FDA classification date
Sep 4, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a risk of false positive results that could lead to unnecessary medical treatment.

Source facts

Action
On 05/30/2025, the firm sent via email an "URGENT MEDICAL DEVICE NOTIFICATION" to customers informing them that, as a result of an internal investigation, it was revealed that samples taken from a frozen negative control panel tested unexpectedly interpreted as equivocal or positive results. Customer are instructed to: . As the issue can alter product performance of the affected lots by reducing specificity and increasing the number of equivocal results, interpret positive results with caution and contact Bio-Rad Technical Support if you observe an unusual rate of false positive results. . Refer to the test limitations, as outlined in Section 9 Limitations of the Procedure, of the IFU 0002043 version 2023/11: Diagnosis of recent infection by T. gondii can only be established on the basis of a combination of clinical and serological data. The result of a single serum sample does not constitute sufficient proof for diagnosis of recent infection . . As noted in Section 8 Interpretation of Results, positive IgM results should be followed by confirmatory steps in accordance with CDC guidelines before initiating or continuing treatment. For questions or assistance, contact Bio-Rad Technical Support at 1-800-224-6723 or by email at TechSupportUSSD-Redmond@Bio-Rad.com
Initiated
2025-05-30
Root cause
Process control
Distribution
U.S. Nationwide distribution in the states of CA, CO, FL, NC and NV.
Recall number
Z-2515-2025
Recall status
Open, Classified
CDRH record ID
214583
FDA classified
2025-09-04
Recall event number
97106

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗