Open original FDA record ↗Recall record
- FDA identifier
- Z-2517-2023
- Company in FDA record
- Philips Ultrasound, Inc.
- FDA classification date
- Sep 6, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Transducer, an ultrasound system accessory, consists of two parts that are bonded together that may come apart due to chassis bonding issue, which may pose risk to users and patients of tissue damage, and electric shock if the built-in safety measures to prevent electric were to also fail.
Source facts
- Action
- On 6/30/23, recall notices were sent to customers who were advised to do the following: 1) Stop usage and/or distribution of affected transducers and secure them in a location where they cannot inadvertently be used. 2) Circulate the notice to all users of this device so that they are aware of the issue. Retain this notice with your system(s) and ensure the letter is in a place likely to be seen/viewed. 3) Complete and return the response form via email to: ultrasound.corrections@philips.com Field Service Engineers will visit your site to provide a replacement device, and collect the affected product. Contact Customer Care Solutions Center if you need further information or support concerning this issue at 1-800-722-9377. On 7/24/23 additional recall notices were distributed to customers informing them that additional devices were found to be affected.
- Initiated
- 2023-06-30
- Root cause
- Process control
- Distribution
- Worldwide - US Nationwide distribution including in the states of FL, MN, AZ, NC, CA, MO, OH, WV, AL, NH, IN, IL, VT, PA, NY, NE, TX, WI, MI, KY, NV, MA, SD, LA, CO, TN, AK, GA, WA, UT, NM, ME, AR, SC and the countries of Italy, China, Netherlands, New Zealand, Hungary, Australia, France, Thailand, C¿te D'Ivoire, Germany, Canada, Mexico, Spain, Saudi Arabia, Uzbekistan, Greece, Yemen.
- Recall number
- Z-2517-2023
- Recall status
- Completed
- CDRH record ID
- 202483
- FDA classified
- 2023-09-06
- Recall event number
- 92761