Z-2517-2023: 3D9-3v Transducer, accessory to Philips Ultrasound Systems: EPIQ Elite; Affiniti 30, 50, and 70; ClearVue

Recall record Z-2517-2023 for Philips Ultrasound, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2517-2023
Company in FDA record
Philips Ultrasound, Inc.
FDA classification date
Sep 6, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Transducer, an ultrasound system accessory, consists of two parts that are bonded together that may come apart due to chassis bonding issue, which may pose risk to users and patients of tissue damage, and electric shock if the built-in safety measures to prevent electric were to also fail.

Source facts

Action
On 6/30/23, recall notices were sent to customers who were advised to do the following: 1) Stop usage and/or distribution of affected transducers and secure them in a location where they cannot inadvertently be used. 2) Circulate the notice to all users of this device so that they are aware of the issue. Retain this notice with your system(s) and ensure the letter is in a place likely to be seen/viewed. 3) Complete and return the response form via email to: ultrasound.corrections@philips.com Field Service Engineers will visit your site to provide a replacement device, and collect the affected product. Contact Customer Care Solutions Center if you need further information or support concerning this issue at 1-800-722-9377. On 7/24/23 additional recall notices were distributed to customers informing them that additional devices were found to be affected.
Initiated
2023-06-30
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution including in the states of FL, MN, AZ, NC, CA, MO, OH, WV, AL, NH, IN, IL, VT, PA, NY, NE, TX, WI, MI, KY, NV, MA, SD, LA, CO, TN, AK, GA, WA, UT, NM, ME, AR, SC and the countries of Italy, China, Netherlands, New Zealand, Hungary, Australia, France, Thailand, C¿te D'Ivoire, Germany, Canada, Mexico, Spain, Saudi Arabia, Uzbekistan, Greece, Yemen.
Recall number
Z-2517-2023
Recall status
Completed
CDRH record ID
202483
FDA classified
2023-09-06
Recall event number
92761

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗