Z-2517-2026: Splashwire Hydrophilic Guide Wire: REF: MSWSTDA35150 UDI-DI codes: 00884450529645 00884450529669 00884450

Recall record Z-2517-2026 for Merit Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2517-2026
Company in FDA record
Merit Medical Systems, Inc.
FDA classification date
Jun 18, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to unsealed portions of pouches.

Source facts

Action
On 05/08/2026, the firm sent via FedEx Express an "U.S. URGENT MEDICAL DEVICE RECALL NOTICE" to inform customers that internal review identified that certain devices manufactured during a defined time period may have an incomplete seal in the pouch of the Merit SplashWire Hydrophilic Steerable Guide Wire. Customer are instructed to: 1. Immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within their facility. 2. A picture of the SplashWire Hydrophilic Steerable Guide Wire Hydrophilic Steerable Guide Wire is attached to help them identify the product. 3. Immediately discontinue the use of the affected lots of the SplashWire Hydrophilic Steerable Guide Wire. 4. Ensure that applicable personnel within their organization are made aware of this recall. 5. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. 6. Fill out, scan and email the completed Customer Response Form to Customer Service at RESPONSE@merit.com within 10 business days. All affected product shipped to them must be accounted for on the CRF. 7. Immediately return all affected lots in their possession to Merit, per the instructions in the attached CRF. 8. When requesting replacement for the SplashWire Hydrophilic Steerable Guide Wire please utilize the table for the replacement part number and contact RESPONSE@merit.com, when they provide their completed Customer Response Form. For questions - contact your Merit Sales Representative or Merit Customer Service via email at RESPONSE@merit.com or via phone at +1 800 356 3748 | Hours: 6 am to 6 pm MST | Mon-Fri.
Initiated
2026-05-08
Root cause
Process design
Distribution
Worldwide - U.S. Nationwide distribution in the states and territory AL, AZ, CA, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and Puerto Rico. The countries of Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece, Guatemala, Ireland, Italy, Kazakhstan, Kenya, Netherlands, Norway, Portugal, Spain, Sweden, and Switzerland.
Recall number
Z-2517-2026
Recall status
Open, Classified
CDRH record ID
219884
FDA classified
2026-06-18
Recall event number
98818

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗