Z-2518-2025: Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions 11.1 and 11.1.1 Product Desc

Recall record Z-2518-2025 for Merge Healthcare, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2518-2025
Company in FDA record
Merge Healthcare, Inc.
FDA classification date
Sep 4, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain units trigger an error message on the Merge Hemo Client (Pressure Sensor Defective) during non-invasive blood pressure (NIBP) measurements, specifically when the initial inflation pressure is set in the high-pressure range (>240 mmHg). The ARGUS PB-3000 is a vital signs monitoring instrument intended to be integrated into the Merge Hemo System. No other components of the Merge Hemo System are affected.

Source facts

Action
Merge Healthcare issued a Field Safety Corrective Action notice to its consignees on 08/12/2025 via email. The notice explained the issue with the device, potential risk to patients, and requested the following actions: Immediate actions to reduce risk to patients For customers using affected versions of Merge Hemo, the following action should be taken to mitigate the risk of non-invasive blood pressure (NIBP) becoming non-functional in the high-pressure range (>240 mmHg): " Ensure an alternative method for determining NIBP, such as a sphygmomanometer or other NIBP measuring device, is available during use of the Merge Hemo system. Actions being taken by Merge Merge Healthcare will contact you to arrange a no-charge inspection of your ARGUS PB-3000 device(s). If the device is affected by this issue, Merge Healthcare will provide a replacement ARGUS PB-3000 and send the affected device to Schiller AG. Please forward this notice to any individuals within your organization that may need to be aware of this notification. Customers should alert any other affiliated parties that may be affected by this Field Safety Corrective Action. Customers experiencing the reported issue should contact Merge Healthcare Support at 1-877-741-5369.
Initiated
2025-08-11
Root cause
Process control
Distribution
US Nationwide distribution including in the states of Arizona, Arkansas, California, Florida, Louisiana, Mississippi, North Carolina, North Dakota, Oklahoma, Pennsylvania, Georgia, Illinois, Kentucky, Tennessee, Texas, Washington, and Wisconsin.
Recall number
Z-2518-2025
Recall status
Open, Classified
CDRH record ID
215580
FDA classified
2025-09-04
Recall event number
97429

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗