Skip to content

Public FDA record

Z-2519-2023: M/L Taper Hip Prosthesis Extended Offset, Reduced Neck Length, Size 4 (Item Number 00771100640) and Size

Recall record Z-2519-2023 for Zimmer, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2519-2023
Company in FDA record
Zimmer, Inc.
FDA classification date
Sep 7, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The outer package labeling and product etch are for a Size 6, however, the implant is a Size 4, and vice versa.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 7/26/23 was sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. You are receiving this notice because you have implanted one of the impacted stems based on our records. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. It is anticipated that the fit of the stem would have been highly detectable, for example a physical size 4 stem sitting lower than expected in a bone prepared for size 6, or vice versa and the stem sitting proud. During surgery, the available offsets in femoral head implants may have provided the necessary flexibility to appropriate
Initiated
2023-07-26
Root cause
Mixed-up of materials/components
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, GA, MI, NJ, PA, TN, and WI. The countries of Brazil, Canada, Netherlands.
Recall number
Z-2519-2023
Recall status
Open, Classified
CDRH record ID
202821
FDA classified
2023-09-07
Recall event number
92889

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗