Z-2521-2018: CORAIL Cementless Femoral Stem HA Coated Standard Size 11, Catalog No. 3L92511

Recall record Z-2521-2018 for DePuy Orthopaedics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2521-2018
Company in FDA record
DePuy Orthopaedics, Inc.
FDA record creation date
Oct 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Two lots of femoral stems were labeled as Size 11. However, they are actually Size 12. Potential adverse effects include poor joint mechanics, improper fit, loosening, and additional surgical procedure to correct. It may also result in an intraoperative fracture which could represent a surgical delay and additional procedures.

Source facts

Action
On June 21, 2018, the firm issued an Urgent Information: US Recall Notice via email to affected customers. Medical professionals were asked to do the following: " Cease using the affected components immediately. " Immediately return unused affected units: Medical facilities are to determine if any of the recalled components are still on hand, and return affected devices immediately to their DePuy Sales Consultant or return them to DePuy for credit following normal purchasing procedures. Reconciliation forms should be faxed to 574-371-4939. " Reconciliation Form: Complete the Reconciliation Form and return to your DePuy Synthes Sales Consultant or fax to 574-371-4939 within five (5) days of this notice. " Records: Retain a copy of the completed Reconciliation Form in your files along with this notice. " Additional Notifications: Notify surgeon users at your facility by providing them with a copy of this notice to ensure surgeon users are aware of this recall notice. Forward this notice to others in your facility that need to be informed. " Maintain a copy of this notice with the affected devices. For questions about device recall information provided, please contact Kim Long, Senior Recall Coordinator, at 574-371-4917 or klong16@its.jnj.com. U.S. Distributors and Sales Offices received a separate notice with specific instructions on how to identify, remove, and return the affected devices.
Initiated
2018-06-21
Root cause
Process control
Terminated
2020-04-06
Distribution
Accounts in AL, CA, CO, GA< IL, LA, ME, NC, NH, NY, OR, TX, UT, and WA. Foreign distribution to Belgium, Croatia, Cyprus, Denmark, Germany, Indonesia, Korea, Sweden, and Tunisia.
Recall number
Z-2521-2018
Recall status
Terminated
CDRH record ID
164940
FDA record created
2018-10-11
Recall event number
80240

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗