Two issues were identified with the stationary x-ray system. 1. Some position illustrations in the patient exam interface show the incorrect anatomical position for the standard of care. When the image is acquired with the wrong reference position, clinical diagnostic requirements cannot be met.T2. Where image stitching is used, the left and right tag information may be lost. When these problems are encountered, an additional exposure is needed and the patient will receive more radiation.
Source facts
Action
The firm sent a Customer Notification Letter by email on February 27, 2020. The letter included identification of the serial numbers of devices affected by this issue. The firm's direct customers were instructed to identify users of the affected systems which have been installed. Customers were informed that they should not refer to illustrations showing the incorrect anatomical position for the standard of care, and that the impacted systems will be immediately remedied with a software hotfix on R002.19.0-4.8648-690113 software version via field software update.
Customers with questions were instructed to call (855) 221-1552 (toll free) or (832) 699-2799 (direct).
Initiated
2020-02-27
Root cause
Software design
Terminated
2021-05-28
Distribution
US Nationwide distribution including in the state of Texas.
Recall number
Z-2521-2020
Recall status
Terminated
CDRH record ID
180256
FDA record created
2020-07-08
Recall event number
85196