Z-2521-2025: MI Cancer Seek REF MSC000 UDI Code: (01)00860008613325(21)H5HJHDSXF MI Cancer Seek is a single-site assa

Recall record Z-2521-2025 for CARIS LIFE SCIENCES. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2521-2025
Company in FDA record
CARIS LIFE SCIENCES
FDA classification date
Sep 4, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to an incorrect test results provided that indicated the incorrect drug therapy recommendation

Source facts

Action
On 07/14/2025, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter its customer that inform them that an incorrect CDx report was issued on April 14, 2025 with the wild-type KRAS/NRAS results with VECTIBIX (panitumumab) listed as a therapeutic option. The potential cause of the issue occurred due to two unanticipated events occurring in succession within the operational aspect of the report generation software processes of MI Cancer Seek. These two events in succession resulted in a failing file. Customer was provided with an amended report and instructed to: -Review the report paying careful attention to the amended CDx report with the header "MI Cancer Seek" which changes the wild-type KRAS/NRAS result Detected to wild-type KRAS/NRAS result to Not Detected and the KRAS G12C variant listed in the Other Alterations and Biomarkers Identified section towards the bottom of the page. The new CDx report also includes the banner, Findings did not yield any biomarker results with Companion Diagnostic (CDx) claims . For questions, contact Regulatory Affairs at 351-220-3883 Monday thru Friday 8:00 AM to 5:00 PM Mountain Standard Time or email Pcarter@carisls.com. In addition Caris Customer Support +1-888-979-8669
Initiated
2025-06-17
Root cause
Software design
Distribution
U.S. Nationwide distribution in the state of NY.
Recall number
Z-2521-2025
Recall status
Open, Classified
CDRH record ID
214979
FDA classified
2025-09-04
Recall event number
97258

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗