Open original FDA record ↗Recall record
- FDA identifier
- Z-2522-2018
- Company in FDA record
- Abbott Laboratories, Inc.
- FDA record creation date
- Jul 20, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Abbott is advising customers that a small number of CardioMEMS(R) Hospital Electronics Systems (Model CM3000) and Patient Electronics Systems (Model CM1100) may deliver a system error, known as Error 5. While this error message is intended to present if the electronics system exceeds a certain temperature, these units may deliver a false Error 5 message due to an incorrectly configured component within the device electronics.
Source facts
- Action
- The firm notified the hospital and clinics of the recall on 06/14/2018 and the patients on 06/15/2018 by letter. The letter explained that the units may deliver a system error, known as Error 5. While this error message is intended to present if the electronics system exceeds a certain temperature, these units may deliver a false Error 5 message due to an incorrectly configured component within the device electronics. The firm is requesting return of the units. Patients may contact Abbott Remote Care Technical Support at 1-844-692-6367 for assistance and further information.
- Initiated
- 2018-06-14
- Root cause
- Process control
- Terminated
- 2021-09-08
- Distribution
- US, Canada, Netherlands, Germany, Great Britain, France, Ireland, Belgium
- Recall number
- Z-2522-2018
- Recall status
- Terminated
- CDRH record ID
- 164914
- FDA record created
- 2018-07-20
- Recall event number
- 80230