Skip to content

Public FDA record

Z-2522-2018: CardioMEMS HF System Hospital and Patient Electronics Units: (a) Patient Electronics System, Model CM1100

Recall record Z-2522-2018 for Abbott Laboratories, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2522-2018
Company in FDA record
Abbott Laboratories, Inc.
FDA record creation date
Jul 20, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Abbott is advising customers that a small number of CardioMEMS(R) Hospital Electronics Systems (Model CM3000) and Patient Electronics Systems (Model CM1100) may deliver a system error, known as Error 5. While this error message is intended to present if the electronics system exceeds a certain temperature, these units may deliver a false Error 5 message due to an incorrectly configured component within the device electronics.

Source facts

Action
The firm notified the hospital and clinics of the recall on 06/14/2018 and the patients on 06/15/2018 by letter. The letter explained that the units may deliver a system error, known as Error 5. While this error message is intended to present if the electronics system exceeds a certain temperature, these units may deliver a false Error 5 message due to an incorrectly configured component within the device electronics. The firm is requesting return of the units. Patients may contact Abbott Remote Care Technical Support at 1-844-692-6367 for assistance and further information.
Initiated
2018-06-14
Root cause
Process control
Terminated
2021-09-08
Distribution
US, Canada, Netherlands, Germany, Great Britain, France, Ireland, Belgium
Recall number
Z-2522-2018
Recall status
Terminated
CDRH record ID
164914
FDA record created
2018-07-20
Recall event number
80230

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗