Skip to content

Public FDA record

Z-2523-2023: Reusable distal attachments Models: MH-462, MH-463, MH-464, MH-465, MH-466, MH-483, MH-587, MH-588(US), M

Recall record Z-2523-2023 for Aomori Olympus Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2523-2023
Company in FDA record
Aomori Olympus Co., Ltd.
FDA classification date
Sep 7, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002

Source facts

Action
Olympus issued Urgent Medical Device Recall Letter dated July 27, 2023 to ENDOSCOPY DEPARTMENT, RISK MANAGEMENT. Letter states reason for recall, health risk and action to take: Immediately assess any product you have to identify the product number with affected lot listed in this communication, cease use and quarantine any affected products. The image below depicts the area where the lot number is identified. The lot# is on the Carton box and Pack. Please refer to Attachment 2 for the identification of the subject products. 2. Call your Olympus customer service representative at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility for your affected product. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the file (recall) number: 0429 4. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. Olympus requests that you report complaints, including patient physical condition suspected to be affected by DEHP, to Olympus. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). If you require additional information, please contact Tara Safi, Field Corrective Actions Administrator at Tara.Safi@Olympus.com.
Initiated
2023-07-27
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution.
Recall number
Z-2523-2023
Recall status
Open, Classified
CDRH record ID
202627
FDA classified
2023-09-07
Recall event number
92784

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗