Open original FDA record ↗Recall record
- FDA identifier
- Z-2524-2026
- Company in FDA record
- TMJ Solutions Inc
- FDA classification date
- Jun 22, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Custom temporomandibular joint implant may have contained incorrect patient-specific components that led to a poor fit of the implant, leading to a right-sided open bite.
Source facts
- Action
- On 2/11/2026, the customer was contacted and verbally informed of the recall. On 4/23/2026, a recall notice was sent to the customer asking them to take the following action: 1) Inform all appropriate members of your organization who need to be made aware of this action. 2) The affected product is already implanted. Following explantation, during revision surgery, the explanted device shall be returned to the firm's sales representative. 3) Complete and return the Business Reply Form (BRF) via email to cmf-pfa@stryker.com Please contact the firm's quality department cmf-pfa@stryker.com with questions or concerns.
- Initiated
- 2026-02-11
- Root cause
- Process control
- Distribution
- US Nationwide distribution in the state of MA.
- Recall number
- Z-2524-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218948
- FDA classified
- 2026-06-22
- Recall event number
- 98509