Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure
Source facts
Action
Siemens issued on May 25, 2018 an URGENT: MEDICAL DEVICE CORRECTION customer notification to affected customers and advised a hardware update will be initiated . Users advised of the problem, potential risks and action to be taken. Siemens will correct the error by replacing all affected cooling units with Update Instruction AX076/17/S. Users may continue to use the Artis system while awaiting implementation of the corrective action.
Initiated
2018-05-25
Root cause
Process control
Terminated
2019-05-16
Distribution
HI , IL, MI, NC, NY, VA, WI
Recall number
Z-2525-2018
Recall status
Terminated
CDRH record ID
165565
FDA record created
2018-07-20
Recall event number
80476