- FDA identifier
- Z-2525-2023
- Company in FDA record
- Hamilton Medical, Inc.
- FDA classification date
- Sep 20, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Source facts
- Action
- On July 26, 2023, Hamilton Medical issued Urgent Medical Device Correction via E-Mail. Hamilton Medical asked consignees to take the following actions 1. Check for affected devices in your facility. Get in contact with your local distribution partner and get the devices serviced with high priority. 2. Please sign the enclosed/attached sheet to confirm that, as an End- Customer, you have received and duly noted this Medical Device Correction.
- Initiated
- 2023-07-26
- Root cause
- Nonconforming Material/Component
- Distribution
- Distribution US nationwide, including Puerto Rico. OUS: Not provided
- Recall number
- Z-2525-2023
- Recall status
- Open, Classified
- CDRH record ID
- 202790
- FDA classified
- 2023-09-20
- Recall event number
- 92873