Z-2525-2026: VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001

Recall record Z-2525-2026 for Ventec Life Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2525-2026
Company in FDA record
Ventec Life Systems, Inc.
FDA classification date
Jun 23, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.

Source facts

Action
On 5/8/2026, recall notices were mailed to customers who were asked to do the following: 1) Immediately discontinue use of affected devices. 2) Inspect current stock, quarantine/segregate any unused affected products, remove affected devices from clinical service and segregate to prevent further use. 3) Use an unaffected ventilator for patient support. 4) If you have distributed or transferred affected devices to other facilities, caregivers, or end users, you must promptly notify them of this recall and provide a copy of this notification. 5) Confirm receipt of this notification letter by returning the enclosed Customer Response Form via email to Ventec-Recall@reacthealth.com Upon receipt of the Customer Response Form, firm will arrange for repair of existing device(s). Return the affected product to React Health If you have any questions, please contact firm at 844-698-6276.
Initiated
2026-05-08
Root cause
Vendor change control
Distribution
US Nationwide distribution in the state of UT.
Recall number
Z-2525-2026
Recall status
Open, Classified
CDRH record ID
220084
FDA classified
2026-06-23
Recall event number
98888

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗