Open original FDA record ↗Recall record
- FDA identifier
- Z-2526-2026
- Company in FDA record
- PFM MEDICAL INC.
- FDA classification date
- Jun 23, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.
Source facts
- Action
- On 4/21/2026, recall notices were emailed to customers who were asked to do the following: 1. Inform all customers (hospitals and clinics) and device users (physicians) who have received or used affected batches. This notification should also be shared with any organizations where the potentially affected devices have been transferred. 2. Return unused devices from the affected batches to PFM Medical, Inc. 3. Complete the enclosed Recall Response Form and email back to recall@pfmmedicalusa.com
- Initiated
- 2026-04-21
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the states of NY, MD.
- Recall number
- Z-2526-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220251
- FDA classified
- 2026-06-23
- Recall event number
- 98942