Open original FDA record ↗Recall record
- FDA identifier
- Z-2527-2018
- Company in FDA record
- GE Medical Systems, LLC
- FDA record creation date
- Jul 23, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential safety issue with the table transfer release of the MR Surgical Suite System which may cause an unexpected loss of function of the patient transfer mechanism that allows the patient to be moved from the GE MR Surgical Table to the Maquet Magnus Surgical Table.
Source facts
- Action
- GE Healthcare notified customers on about 03/30/2018 via URGENT MEDICAL DEVICE CORRECTION letter. The letter informed customers that the MR Surgical Suited System could still remain in use, it referred them to the GE MR Surgical Suite Operator's Manual for detailed descriptions of the various safety interlocks and operating mechanisms, encouraged users to monitor for any issues releasing the transfer board from the GE MR Surgical Table, and instructed users to test the patient transfer features as part of a quality check prior to each use. The letter also informed customers that a GE Healthcare representative will contact customers to ensure they have the latest operator manual at their site, update preventative maintenance documentation to include a procedure that focuses specifically on this equipment, and GE Healthcare representative will contact customers for scheduling a visit at which time GEHC will conduct this preventative maintenance procedure for their device and install new hardware. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your the Service Representative.
- Initiated
- 2018-03-30
- Root cause
- Nonconforming Material/Component
- Terminated
- 2024-03-29
- Distribution
- US Nationwide distribution to AZ, CA, CO, FL, IL, KS, MI, MN, NY, OH, PA, TX, UT, VA, WI. Worldwide distribution to Austria, Brazil, China, Czech Republic, France, Germany, India, Italy, Japan, Korea, Kuwait, South Africa, Turkey, Turkey, United Arab Emirates, United Kingdom.
- Recall number
- Z-2527-2018
- Recall status
- Terminated
- CDRH record ID
- 165701
- FDA record created
- 2018-07-23
- Recall event number
- 80287