Skip to content

Public FDA record

Z-2527-2023: Hamilton-T1, REF: 161009, Intensive Care and Transport Ventilator

Recall record Z-2527-2023 for Hamilton Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2527-2023
Company in FDA record
Hamilton Medical, Inc.
FDA classification date
Sep 20, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Source facts

Action
On July 26, 2023, Hamilton Medical issued Urgent Medical Device Correction via E-Mail. Hamilton Medical asked consignees to take the following actions 1. Check for affected devices in your facility. Get in contact with your local distribution partner and get the devices serviced with high priority. 2. Please sign the enclosed/attached sheet to confirm that, as an End- Customer, you have received and duly noted this Medical Device Correction.
Initiated
2023-07-26
Root cause
Nonconforming Material/Component
Distribution
Distribution US nationwide, including Puerto Rico. OUS: Not provided
Recall number
Z-2527-2023
Recall status
Open, Classified
CDRH record ID
202792
FDA classified
2023-09-20
Recall event number
92873

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗