Philips sent an Urgent Medical Device Correction letter dated December 17, 2010, to all affected consignees. Philips will replace defective system boards in Hewlett Packard model rp5700 PCs sold with IntelliVue Information Centers to correct the problem. Philips Healthcare representatives will inspect all affected units, and any unit identified to have a defective system board will have its system board replaced at no charge to the customer.
Philips is asking the following of customers in the Action to be Taken by Customer/User section of the FSN:
Observe your system for any of the symptoms described in the Problem Description section above. If symptoms are noted, please contact your local Philips Healthcare Customer Service representative or the Customer Care Solutions Center at (800) 722-9377 # 3, # 1. Do not rely exclusively on the audible alarm system for patient monitoring. Remember that the most reliable method of patient monitoring combines close personal surveillance with correct operation of monitoring equipment.
For questions regarding this recall call 978-659-3000.
Initiated
2010-12-17
Root cause
Component design/selection
Terminated
2017-08-01
Distribution
Worldwide Distribution - US (Nationwide)
OUS:
CANADA
ARGENTINA
AUSTRALIA
AUSTRIA
BAHRAIN
BELGIUM
BRAZIL
CHILE
CHINA
COLOMBIA
CZECH REPUBLIC
FINLAND
FRANCE
GERMANY
GREECE
INDIA
INDONESIA
IRELAND
ITALY
JAPAN
KENYA
KOREA, REPUBLIC OF
NETHERLANDS
PHILIPPINES
POLAND
SPAIN
SWEDEN
SWITZERLAND
THAILAND
TURKEY
UNITED KINGDOM
Recall number
Z-2528-2017
Recall status
Terminated
CDRH record ID
156055
FDA classified
2017-06-13
Recall event number
77437