Z-2529-2024: Leica HistoCore Arcadia H, REF: 14039354100 and 14039357258, Paraffin Embedding Station.

Recall record Z-2529-2024 for LEICA BIOSYSTEMS NUSSLOCH GMBH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2529-2024
Company in FDA record
LEICA BIOSYSTEMS NUSSLOCH GMBH
FDA classification date
Aug 7, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An issue with safe usage of the device was identified whereby toxic smoke and internal fire developed resulting from a problem associated with the sealing of the paraffin tank.

Source facts

Action
The recalling firm issued letters dated 4/30/2024 via email on 5/10/2024. The letter provided the model and serial number of the devices involved, a description of the problem, and immediate actions to be taken. The consignee is to check to see if they have one of the affected instruments and determine whether the Arcadia H is in active use in the facility. If it is, the consignee is to discontinue the use of it immediately and disconnect it from main voltage. If it has not been decommissioned at the time of receipt of the letter, the consignee is to reach out to their sales representative. The sales representative will work with the consignee to determine the disposition of the device according to their local regulations. The consignee is requested to pass the notification letter to users of the device and to all within their organization who need to know. The consignee is to confirm receipt of the letter within 5 days by completing the attached acknowledgement form and returning a scanned copy via email.
Initiated
2024-05-10
Root cause
Device Design
Distribution
US Nationwide, including government distribution.
Recall number
Z-2529-2024
Recall status
Open, Classified
CDRH record ID
208257
FDA classified
2024-08-07
Recall event number
94776

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗