Z-2529-2025: Artis Pheno. Image-Intensified Flouroscopic X-Ray System.

Recall record Z-2529-2025 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2529-2025
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Sep 5, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Limited system movements after startup .

Source facts

Action
Consignees were mailed an URGENT: MEDICAL DEVICE CORRECTION notification dated 8/12/25. The notification informs consignees that the recall issue will be corrected via release of an Update Instruction AX032/25/S by appointment, which can be arranged by contacting 1-800-888-7436. Consignees are to disseminate the recall notification to all users of affected devices within their organization and to where product may be further distributed. Consignees are to return the completed recall response form via email to recallsandrefusals.team@siemens-healthineers.com or by certified mail.
Initiated
2025-08-12
Root cause
Software design
Distribution
US Nationwide distribution in the states of AL, AZ, FL, KY, MA, MI, NY, OR, A, TN, TX & UT.
Recall number
Z-2529-2025
Recall status
Open, Classified
CDRH record ID
215610
FDA classified
2025-09-05
Recall event number
97444

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗