Z-2530-2020: PHANTOM CS QUICK START DISTRACTION SCREW, 16 MM, STERILE, 5/BOX - Product Usage: intended to be used for

Recall record Z-2530-2020 for TeDan Surgical Innovations LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2530-2020
Company in FDA record
TeDan Surgical Innovations LLC
FDA record creation date
Jul 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Lack of pouch seal. Product is labelled as sterile.

Source facts

Action
On June 4, 2020, the firm sent an "URGENT: MEDICAL DEVICE FIELD SAFETY NOTICE" along a "Field Safety Notice Acknowledgement and Receipt Form Response Required" to customers via email informing them the Recaling Firm has received two (2) complaints associated with the packaging issue where there was no seal on the screw pouch opposite the chevron manufacturing pouch seal. The Recalling Firm is instructing customer to: -immediate visual inspection of the pouch be conducted (Letter A shown in the red outline opposite the chevron seal [Letter B] in Figure 1) and observe for an open seal (example photographs are included in the Notification). Any product exhibiting an open seal is to be quarantined, and then returned to the Recalling Firm. Customers are to contact their authorized distributor for RMA and shipping instructions. The Recalling Firm requests the following actions be performed to visually inspect the product packaging: 1. Carefully open the Distraction Screw box only on one (1) end, ensuring box integrity. 2. Remove the five (5) sterile pouches. 3. Visually inspect the seal zone (indicated as letter A in Figure 1) opposite the chevron seal (Letter B) for each sterile pouch. 4. IF the pouch is not sealed, DO NOT USE THE DEVICE. Quarantine the device and contact your authorized distributor who will assist you with shipping instructions and replacements. See contact information below. 5. IF the pouch is sealed, is not opened, the product remains sterile and may be used. 6. Complete and return to TSI the enclosed FIELD SAFETY NOTICE ACKNOWLEDGMENT via fax or e-mail. Please complete as soon as possible but no later than July 6, 2020. Contact Information for Returns - RMA: Account Management Telephone 713-726-0886 Email Address acctmgmt@tedansurgical.com Fax 713-726-0846 For further questions, contact Lynne Davies, ldavies@tedansurgical.com or call 713-554-2130
Initiated
2020-05-15
Root cause
Packaging change control
Terminated
2022-09-08
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of: AZ, CA, FL, GA, HI, IL, MA, MI, NC, NH, NJ, NY, NV, PA, PR, TN, TX, and WA . The countries of Australia, Canada, France, Philippines, Taiwan, U.A.E., and United Kingdom.
Recall number
Z-2530-2020
Recall status
Terminated
CDRH record ID
181860
FDA record created
2020-07-10
Recall event number
85830

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗