Open original FDA record ↗Recall record
- FDA identifier
- Z-2531-2018
- Company in FDA record
- Brewer Company, LLC
- FDA record creation date
- Jul 24, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
One lot, the wire harness was over-crimped on some of the connectors that go to the two vertical side switches on the exam table.
Source facts
- Action
- The Brewer Company notified customers, via "MEDICAL DEVICE CORRECTION" letter, on about 03/21/2018. Instructions included to cease distribution of the product, coordinate with Brewer to schedule rework of the cable in affected tables as quickly as possible, and to notify Brewer of customers if the product was further distributed to coordinate rework of those tables. Additionally secondary controls were provided that the operator must press and continually hold the foot control pedal to operate the table, the foot control includes a stop button, depressing any pedal on the foot control will stop the table's automatic movement, and the manual instructs the operator to remove any obstructions from under the exam table prior to movement. Questions can be directed to (888) 273-9371.
- Initiated
- 2018-03-29
- Root cause
- Nonconforming Material/Component
- Terminated
- 2019-08-29
- Distribution
- Devices distributed in WI, TX, MN and IL.
- Recall number
- Z-2531-2018
- Recall status
- Terminated
- CDRH record ID
- 164734
- FDA record created
- 2018-07-24
- Recall event number
- 80164