Z-2531-2020: Philips M3015A Microstream CO2 Extension Model No. 862393 -used with an lntelliVue X3 Patient Monitor (M

Recall record Z-2531-2020 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2531-2020
Company in FDA record
Philips North America, LLC
FDA record creation date
Jul 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Affected M3015A extensions are incompatible with the lntelliVue X3 Patient Monitor (Model No. 867030) and lntelliVue MMX multi-measurement server (Model No. 867036).If they are connected to a host monitor, communication with the X3 or MMX could be disrupted, resulting in a loss of monitoring on the host monitor.

Source facts

Action
Philips issued Customer Information Medical Device Correction (CIL86201894B) via priority mail on 6/8/20. Philips Market Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take: To check if this issue affects your M3015A, compare the serial number of your device with the affected serial numbers identified; Use affected M3015A extensions only together with an lntelliVue X2 or MMS.; Do not combine an affected M3015A with an lntelliVue X3 or MMX; Philips Healthcare is distributing with this letter an addendum to the Instructions for Use for patient monitors that include directions for the use of the M3015A (MX100/X3, MX400-800 and MX750/MX850). If you need any further information or support concerning this issue, please contact your local Philips representative or Philips Customer Support at 1-800-722-9377.
Initiated
2020-06-08
Root cause
Device Design
Terminated
2023-01-09
Distribution
Nationwide Foreign: Angola Angola Argentina Australia Austria Bahrain Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana Germany Ghana Gibraltar Greece Greenland Guadeloupe Guatemala Hong Kong Hungary Iceland India Indonesia Iran Ireland Israel Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Kuwait Latvia Lebanon Lithuania Luxembourg Malaysia Martinique Mexico Morocco Mozambique Namibia Netherlands New Caledonia New Zealand Nicaragua Norway Oman Palestine, State of Paraguay Peru Philippines Poland Portugal Qatar R¿union Russian Federation Saint Pierre and Miquelon Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Turkey Ukraine United Arab Emirates United Kingdom
Recall number
Z-2531-2020
Recall status
Terminated
CDRH record ID
181796
FDA record created
2020-07-10
Recall event number
85799

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗