Open original FDA record ↗Recall record
- FDA identifier
- Z-2532-2020
- Company in FDA record
- Bound Tree Medical
- FDA record creation date
- Jul 10, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Kit contained an incorrect catheter-tip syringe not compatible with the luer lock connector of the included airway. As a result, may be unable to inflate the balloons of the supraglottic airway to create a secure seal in the patient s airway as intended by the kit
Source facts
- Action
- The firm, Bound Tree, issued "URGENT NOTICE OF RECALL" letter by mail on May 14, 2020 and Bound Tree Customer Service and followed with email or telephone calls. Letter states reason for recall and action to take: Please immediately physically review your inventory of this kit and lot number for the affected syringe . Should you determine your inventory contains affected product, please quarantine affected kits immediately and contact Boundtree customer care to return product for replacement or credit. Questions pertaining to the recall, you may contact Boundtree customer care at 800-533-0523.
- Initiated
- 2020-05-14
- Root cause
- Process control
- Terminated
- 2023-12-07
- Distribution
- Nationwide Distribution
- Recall number
- Z-2532-2020
- Recall status
- Terminated
- CDRH record ID
- 181863
- FDA record created
- 2020-07-10
- Recall event number
- 85832