Skip to content

Public FDA record

Z-2532-2026: Acumed Hexalob Screw various sizes: REF Numbers: : 3070-27008 2.7mm x 8mm NL Hexalobe Screw; 3070-27009 2

Recall record Z-2532-2026 for Acumed LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2532-2026
Company in FDA record
Acumed LLC
FDA classification date
Jun 23, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to manufacturing error, hexalobe screws used with wrist fixation system may break during use.

Source facts

Action
On May 28, 2026 Acumed issued a Urgent Medical Device Recall Notification to affected consignees via phone calls and email. Acumed asked consignees to take the following actions: 1)IMMEDIATELY stop using and/or distributing the product and complete the following actions: 2) Perform a physical count of your inventory. Identify any inventory of the batch/lot numbers listed in the table above. 3) Remove the product segregated in step 1, such that they are taken out of service (i.e., quarantined). 3) Verbally acknowledge (then respond via email) to Kathryn.jayne@acumed.net the recall notice. 4) You will be contacted by Field Inventory department who will directly remove the screws and conduct a count reconciliation. 5) This notice needs to be passed on to all those who need to be aware within your organization or facilities where the affected devices have been transferred.
Initiated
2026-05-28
Root cause
Device Design
Distribution
US Nationwide distribution in the states of AR, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MS, NC, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA, WI.
Recall number
Z-2532-2026
Recall status
Open, Classified
CDRH record ID
220874
FDA classified
2026-06-23
Recall event number
99123

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗